VT-06 Viscotester RION japan

Update Terakhir
06/04/2020 05.23.26
Minimum Pembelian
1 Unit
Harga
Call

Deskripsi Produk

viscometer, viscotester

Usage
1. Attach rotor to unit and hold unit in the hand or place on dedicated stand.
(Unit should be approximately horizontal in either case.)
2. Insert rotor in sample fluid, turn power on, and select rotor number.
3. Press start button and read indicated viscosity.
The supplied extension rod can reach fluid that is further away. (Only for use with the No.1 and No.2 rotors.)
The VT-06 is designed for quality control applications in the manufacturing process of
industrial products such as petrochemicals, paint, and adhesives, as well as foodstuffs.
Viscosity measurements covering a wide range are possible, such as gear oil used in
construction machinery. Measurement is performed by simply submerging a rotor in the
fluid. The resistance to rotor movement caused by the viscosity (torque) is measured to
obtain direct readings.
Compact and light weight make the unit easily portable and
allow operation with one hand
Can be powered by alkaline batteries, nickel-hydride rechargeable batteries,
or AC adapter
Direct indication of viscosity in millipascal-seconds or decipascal-seconds (SI units)
Dedicated stand for measurement available as option
1 JIS R 3503 : 1994, φ78×103(H
Specifications
Measurement range
Sample fluid capacity
Measurement accuracy
Rotor speed
Power supply
Dimensions and Weight
Supplied accessories
175 (H) × 77 (W) × 40 (D) mm (without protruding parts),
Approx. 260 g (without batteries)
No. 1 rotor (dia. 24 × 53 × 166 mm) SUS304 1
No. 2 rotor (dia. 15 × 1 × 113 mm) SUS304 1
No. 3 rotor (dia. 45 × 47 × 160 mm) SUS304 1
No. 3 Cup (dia. 52.6 × 75 mm) SUS304 1
Extension rod (900 mm300× 3) SUS304 1
IEC LR6 (size AA) alkaline batteries 4
No. 3 rotor: 0.3 to 13 dPa.s (with No. 3 cup)
No. 1 rotor: 3 to 150 dPa.s (with JIS 300 mL beaker*1 )
No. 2 rotor: 100 to 4000 dPa.s (with JIS 300 mL beaker*1 )
No. 1 and No. 2 rotor (with JIS 300 mL beaker*1 ) approx. 300 mL
No. 3 rotor (with No. 3 cup) approx. 150 mL
Clearance between rotor end and cup bottom:
about 15 mm
±10 %±1 digit of indicated value, reproducibility ±5 %
62.5 rpm
IEC LR6 (size AA) alkaline batteries,
nickel-hydride rechargeable batteries, AC adapter VA-05J
Stand
AC adapter
Options
VA-04
VA-05J
Product name Product number